Device recall · February 19, 2020
Prometra II 40mL Pump. Catalog No. 16827 The device is an implantable infusion pump. GTIN-14 00810335020242 recall
Why: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Prometra II 40mL Pump. Catalog No. 16827 The device is an implantable infusion pump. GTIN-14 00810335020242
- Lots / codes
All units with pump firmware version 0.26 manufactured from 2012 until the present time. Serial Numbers 10GC4A24X 10GC4A02X 10GC4A18X 10GC4A22X 10GC4A14X 4MA00044X 4MA00006X 10GC4A12X 10GC4A09X 10GC4A06X 10GC4A21X 4MA00008X 10GC4A01X 10GC4A20X 10GC4A11X 10GC4A16X 10GC4A17X 4MA00045X 4MA00055X 10GC4A08X 4MA00032X 4MA00031X 4MA00010X 4MA00042X
- Quantity
- 6670 (US)
- Company
- Flowonix Medical Inc, Mansfield, MA
- Distribution
- Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
- Initiated
- December 24, 2019 · Voluntary: Firm initiated
- Classified
- February 10, 2020
- Reported
- February 19, 2020
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.