Device recall · February 19, 2020
Prometra II 20mL Pump. Catalog No. 13827 The device is an implantable infusion pump. GTIN-14 00810335020228 recall
Why: A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Prometra II 20mL Pump. Catalog No. 13827 The device is an implantable infusion pump. GTIN-14 00810335020228
- Lots / codes
All units with pump firmware version 0.26 manufactured from 2012 until the present time. Serial Numbers 10HW4A31 10AZ4A69X 10AZ4A07X 10BZ4A16X 10BZ4A61X 10BZ4A41X 10AZ4A48X 10AZ4A50X 10AZ4A23X 10AZ4A36X 10AZ4A42X 10BZ4A37X 10BZ4A64X 10BZ4A12X 10CZ4B09X 10CZ4A95X 10CZ4B22X 10BZ4A18X 10CZ4A79X 10CZ4A86X 10CZ4A96X 10CZ4A88X 10CZ4A80X 10CZ4A85X 10CZ4A81X 10CZ4A82X 10CZ4A78X 10CZ4A91X 10CZ4A94X 10CZ4A90X 10CZ4A77X 10CZ4B28X 10CZ4B05X 10CZ4B15X 10CZ4B21X 10CZ4A87X 10CZ4B32X 10AZ4A41X 10CZ4B23X 10CZ4A93X 10CZ4B01X 10CZ4B17X 10CZ4B04X 10CZ4B30X 10BZ4A55X 10CZ4B08X 10CZ4A76X 10CZ4B33X 10CZ4B20X 10DZ4A01X 10DZ4A58X 10DZ4A19X 10CZ4B36X 10CZ4A84X 10DZ4A18X 10CZ4B24X 10DZ4A76X 10BZ4A30X 10CZ4B26X 10BZ4A09X 10DZ4A51X 10DZ4A93X 10DZ4A32X 10CZ4B03X 10DZ4A89X 10BZ4A11X 10DZ4A30X 10CZ4B25X 10DZ4A65X 10DZ4A05X 10DZ4A98X 10DZ4A08X 10DZ4A42X 10DZ4A77X 10CZ4A92X 10DZ4A35X 10DZ4A80X 10DZ4A92X 10DZ4A88X 10DZ4A07X 10BZ4A39X 10DZ4A73X 10DZ4A53X 10DZ4A28X 10DZ4A84X 10DZ4A45X 10CZ4B14X 10DZ4A40X 10DZ4A72X 10DZ4A64X 10DZ4A79X 10DZ4A60X 10DZ4A82X 10DZ4A37X 10DZ4B39X 10DZ4A54X 10DZ4A67X 10CZ4B29X 10BZ4A59X 10DZ4A20X 10DZ4A38X 10DZ4B04X 10DZ4A36X 10DZ4A91X 10DZ4B37X 10DZ4B06X 10DZ4A34X 10DZ4A25X 10DZ4B24X 10DZ4B49X 10DZ4B38X 10CZ4A89X 10DZ4B05X 10DZ4A15X 10DZ4A43X 10DZ4B54X 10BZ4A25X 10DZ4A66X 10DZ4B53X 10DZ4A29X 10DZ4A23X 10DZ4B08X 10DZ4A21X 10DZ4B40X 10DZ4A83X 10DZ4A17X 10DZ4A62X 10DZ4B19X 10DZ4B46X 10EZ4A36X 10DZ4B16X 10DZ4A87X 10DZ4B25X 10DZ4B43X 10DZ4B10X 10DZ4B61X 10DZ4B50X 10DZ4B07X 10DZ4B48X 10DZ4B26X 10DZ4B29X 10DZ4A27X 10CZ4B27X 10EZ4A02X 10DZ4B22X 10DZ4B15X 10EZ4A46X 10EZ4A27X 10EZ4A01X 10DZ4A47X 10DZ4A81X 10DZ4A33X 10EZ4A30X 10EZ4A26X 10EZ4A51X 10DZ4B03X 10EZ4A37X 10EZ4A13X 10EZ4A39X 10EZ4A35X 10DZ4B45X 10EZ4A28X 10DZ4A10X 10FZ4A34X 10CZ4B35X 10EZ4A09X 10EZ4A43X 10EZ4A05X 10EZ4A17X 10EZ4A16X 10DZ4B21X 10DZ4A94X 10EZ4A19X 10EZ4A45X 10DZ4A44X 10DZ4A68X 10EZ4A25X 10FZ4A29X 10EZ4A32X 10DZ4A48X 10FZ4A24X 10EZ4A18X 10DZ4B55X 10DZ4A06X 10FZ4A02X 10EZ4A57X 10EZ4A24X 10EZ4A58X 10EZ4A06X 10FZ4A38X 10FZ4A03X 10EZ4A29X 10EZ4A08X 10EZ4A55X 10FZ4A36X 10EZ4A20X 10EZ4A31X 10DZ4B18X 10GZ4A40X 10FZ4A18X 10FZ4A39X 10DZ4B42X 10FZ4A31X 10EZ4A47X 10GZ4A44X 10GZ4A58X 10DZ4B51X 10GZ4A27X 10DZ4B30X 10FZ4A37X 10GZ4A01X 10GZ4A43X 10GZ4A38X 10GZ4A35X 10EZ4A48X 10GZ4A03X 10GZ4A08X 10GZ4A37X 10DZ4B41X 10GZ4A02X 10DZ4B31X 10DZ4A55X 10GZ4A76X 10GZ4A66X 10DZ4B20X 10FZ4A15X 10FZ4A01X 10GZ4A34X 10EZ4A10X 10FZ4A22X 10GZ4A73X 10EZ4A53X 10FZ4A21X 10GZ4A94X 10GZ4A06X 10GZ4A61X 10GZ4A78X 10EZ4A60X 10FZ4A13X 10GZ4A96X 10GZ4A04X 10FZ4A28X 10GZ4A11X 10EZ4A34X 10GZ4A59X 10EZ4A11X 10GZ4A05X 10FZ4A04X 10GZ4A18X 10EZ4A49X 10GZ4A54X 10GZ4A92X 10GZ4A90X 10GZ4A57X 10EZ4A40X 10FZ4A40X 10GZ4A63X 10DZ4A49X 10GZ4A41X 10GZ4A29X 10GZ4A83X 10GZ4B54X 10GZ4A70X...
- Quantity
- 6670 (US)
- Company
- Flowonix Medical Inc, Mansfield, MA
- Distribution
- Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.
- Initiated
- December 24, 2019 · Voluntary: Firm initiated
- Classified
- February 10, 2020
- Reported
- February 19, 2020
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.