Drugs A-Z Is my price fair? Price changes Drug markups Drug shortages Recalls Medicare prices Upcoming generics What people paid Guides Share what you paid
Theme
Class IITerminatedZ-0438-2016

Device recall · December 23, 2015

MiniMed 640G recall

Why: The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.

Class II: could cause temporary or reversible health problems; serious harm is remote.

The recall

Product

MiniMed 640G, 3.0mL mmol/L, 3.0mL mg/dL, Model No. MMT-1711 and MMT-1712 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.

Lots / codes

NG1000742H NG1001185H NG1001199H NG1001204H NG1001206H NG1002461H NG1006813H NG1007995H NG1013701H NG1015570H NG1016916H NG1017305H NG1019509H NG1019517H NG1019519H NG1019521H NG1019522H NG1019523H NG1019528H NG1019530H NG1019533H NG1019534H NG1019536H NG1019543H NG1019544H NG1019547H NG1019549H NG1019550H NG1019551H NG1019553H NG1019555H NG1019556H NG1019557H NG1019558H NG1019559H NG1019561H NG1019562H NG1019564H NG1019565H NG1019566H NG1019567H NG1019568H NG1019569H NG1019571H NG1019572H NG1019573H NG1019576H NG1019577H NG1019578H NG1019579H NG1019580H NG1019581H NG1019583H NG1019585H NG1019586H NG1019587H NG1019588H NG1019589H NG1019590H NG1019591H NG1019595H NG1019596H NG1019597H NG1019598H NG1019600H NG1019601H NG1019602H NG1019603H NG1019604H NG1019606H NG1019607H NG1019608H NG1019609H NG1019610H NG1019611H NG1019612H NG1019613H NG1019615H NG1019616H NG1019617H NG1019618H NG1019619H NG1019620H NG1019621H NG1019622H NG1019623H NG1019624H NG1019625H NG1019626H NG1019627H NG1019628H NG1019630H NG1019631H NG1019632H NG1019633H NG1019634H NG1019635H NG1019636H NG1019637H NG1019638H NG1019640H NG1019641H NG1019642H NG1019644H NG1019646H NG1019647H NG1019649H NG1019650H NG1019651H NG1019652H NG1019654H NG1019655H NG1019656H NG1019657H NG1019658H NG1019659H NG1019663H NG1019664H NG1019666H NG1019667H NG1019669H NG1019670H NG1019671H NG1019672H NG1019673H NG1019674H NG1019676H NG1019677H NG1019678H NG1019679H NG1019680H NG1019681H NG1019682H NG1019683H NG1019684H NG1019686H NG1019688H NG1019689H NG1019691H NG1019692H NG1019693H NG1019695H NG1019696H NG1019697H NG1019698H NG1019699H NG1019701H NG1019702H NG1019703H NG1019704H NG1019705H NG1019706H NG1019707H NG1019708H NG1019709H NG1019710H NG1019713H NG1019714H NG1019715H NG1019716H NG1019717H NG1019718H NG1019720H NG1019721H NG1019722H NG1019723H NG1019725H NG1019726H NG1019727H NG1019730H NG1019731H NG1019732H NG1019733H NG1019734H NG1019735H NG1019736H NG1019737H NG1019739H NG1019740H NG1019741H NG1019743H NG1019744H NG1019745H NG1019746H NG1019747H NG1019748H NG1019749H NG1019750H NG1019752H NG1019755H NG1019757H NG1019758H NG1019759H NG1019760H NG1019761H NG1019762H NG1019763H NG1019764H NG1019765H NG1019767H NG1019768H NG1019769H NG1019772H NG1019776H NG1019777H NG1019780H NG1019781H NG1019782H NG1019824H NG1019825H NG1019826H NG1019829H NG1019831H NG1019832H NG1019833H NG1019835H NG1019836H NG1019837H NG1019838H NG1019839H NG1019840H NG1019842H NG1019843H NG1019844H NG1019845H NG1019849H NG1019850H NG1019851H NG1019852H NG1019853H NG1019854H NG1019855H NG1019856H NG1019857H NG1019858H NG1019859H NG1019861H NG1019863H NG1019864H NG1019866H NG1019867H NG1019868H NG1019869H NG1019870H NG1019872H NG1019873H NG1019874H NG1019875H NG1019876H NG1019881H...

Quantity
1052 units
Company
Medtronic MiniMed Inc., Northridge, CA
Distribution
Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.
Initiated
September 25, 2015 · Voluntary: Firm initiated
Classified
December 17, 2015
Reported
December 23, 2015
Ended
February 5, 2016
How notified
Letter

If you have this product

  1. Check the lot number on your package against the lots above. Only those lots are recalled.
  2. Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
  3. Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
  4. Return or dispose of recalled product as the company or your pharmacy advises.

Recalled with it (same recall event)

Questions about this recall

Be kind and stay on topic. No personal health details, links or selling. This is not medical advice.

No comments yet. Questions, tips, what worked for you - start the conversation.

Help the next person

Know what your pharmacy pays.