Device recall · December 23, 2015
MiniMed 620G recall
Why: The pump drive motors may experience a malfunction which would result in a pump error message alarm notifying you that insulin is no longer being delivered. Possible cause is either lack of greasing or deficient greasing in a ball bearing assembly of the motor.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
MiniMed 620G, 3.0mL mg/dL, Model No. MMT-1710 The MiniMed 620G and MiniMed 640G insulin pump systems are indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. In addition, the pump system is indicated for continuous monitoring of glucose levels in the fluid under the skin, and possible low and high blood glucose episodes.
- Lots / codes
NG1019903H NG1019906H NG1019907H NG1019909H NG1019914H NG1019915H NG1019916H NG1019918H NG1019921H NG1019922H NG1019923H NG1019925H NG1019926H NG1019928H NG1019930H NG1019931H NG1019932H NG1019933H NG1019935H NG1019937H NG1019939H NG1019940H NG1019943H NG1019944H NG1019946H NG1019949H NG1019950H NG1019953H NG1019966H NG1019971H NG1019975H NG1019976H NG1019979H NG1019981H NG1019989H NG1019997H NG1019999H NG1020001H NG1020006H NG1020007H NG1020012H NG1020015H NG1020017H NG1020018H NG1020019H NG1020020H NG1020022H NG1020023H NG1020026H NG1020027H NG1020028H NG1020029H NG1020031H
- Quantity
- 53 units
- Company
- Medtronic MiniMed Inc., Northridge, CA
- Distribution
- Internationally Distribution only to the countries of : Australia, Austria, Belgium, Chile, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Italy, Japan, Kenya, Luxembourg, Namibia, Netherlands, Norway, Slovenia, South Africa, Spain, Sweden, Switzerland, United Kingdom.
- Initiated
- September 25, 2015 · Voluntary: Firm initiated
- Classified
- December 17, 2015
- Reported
- December 23, 2015
- Ended
- February 5, 2016
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.