BD Alaris System PC Unit Model 8100 recall
Why: Pump Module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion or interruption of infusion. Delays in an infusion or interruption can cause serious injury or death in high-risk patient populations.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
BD Alaris System PC Unit Model 8100 (Pump Module) and Pump Module Door Assembly Replacement Kits (P/N 49000239, 49000346, 49000438, 49000439) infusion pump.
- Lots / codes
All serial numbers
- Quantity
- 248,519
- Company
- CareFusion 303, Inc., San Diego, CA
- Distribution
- Worldwide: US* (AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY); United Arab Emirates (AE); Australia (AU); Canada (CA); Hungary (HU); Israel (IL); Japan (JP)* Kuwait (KW); New Zealand (NZ); Philippines (PH); Qatar (QA); Saudi Arabia (SA); Singapore (SG); Taiwan (TW); Turkey (TR)*. *There are 159 U.S. government and/or military consignees including two located in Japan (JP) and Turkey (TR).
- Initiated
- August 4, 2020 · Voluntary: Firm initiated
- Classified
- September 10, 2020
- Reported
- September 16, 2020
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.