Device recall · October 5, 2016
Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump recall
Why: A software anomaly with the Alaris Syringe module software version 9.15 may cause an infusion to unexpectedly stop when the Syringe module is transitioning from one rate to another.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Alaris Syringe Module Model 8110 with software version 9.15 The Alaris Syringe Pump module is part of the Alaris System. The syringe pump delivers fluids in a manner similar to current syringe pumps on the market. Up to four Alaris Syringe pump modules can be connected to the Alaris PC unit which is the central programming, monitoring and power supply component for the Alaris System. The syringe pump uses standard, single-use administration sets and syringes with luer-lock connectors.
- Lots / codes
Software version 9.15
- Quantity
- 12,000 units
- Company
- CareFusion 303, Inc., San Diego, CA
- Distribution
- Worldwide Distribution: US(nationwide) and countries of: Canada, Australia and United Arab Emirates.
- Initiated
- August 8, 2016 · Voluntary: Firm initiated
- Classified
- September 23, 2016
- Reported
- October 5, 2016
- Ended
- October 26, 2018
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.