Device recall · October 9, 2019
Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the q recall
Why: Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Accu-Chek Performa II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
- Lots / codes
Serial numbers less than xxx25525056 Updated serial numbers on 10/12/2019: Performa II: Between XXX20000000 and XXX25525056 Updated 10/18/2019: Performa II: from XXX20000000 to XXX25525056
- Quantity
- N/A
- Company
- Roche Diabetes Care, Inc., Indianapolis, IN
- Distribution
- Worldwide distribution - US Nationwide distribution.
- Initiated
- June 12, 2017 · Voluntary: Firm initiated
- Classified
- September 30, 2019
- Reported
- October 9, 2019
- Ended
- October 30, 2020
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.