Device recall · August 12, 2020
Alaris System PC Unit Model 8015. modular infusion pump and monitoring system recall
Why: If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
Alaris System PC Unit Model 8015. modular infusion pump and monitoring system
- Lots / codes
All serial numbers
- Quantity
- 824277
- Company
- CareFusion 303, Inc., San Diego, CA
- Distribution
- U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, BH
- Initiated
- June 30, 2020 · Voluntary: Firm initiated
- Classified
- August 6, 2020
- Reported
- August 12, 2020
- Ended
- April 22, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.