Device recall · August 12, 2020
Alaris SpO2 Module Model 8210 and Model 8220 recall
Why: Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
Alaris SpO2 Module Model 8210 and Model 8220, a modular infusion pump and monitoring system
- Lots / codes
During cleaning, updated cleaning instructions and IUI covers must be used for all serial numbers.
- Quantity
- 7854
- Company
- CareFusion 303, Inc., San Diego, CA
- Distribution
- Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
- Initiated
- June 30, 2020 · Voluntary: Firm initiated
- Classified
- August 6, 2020
- Reported
- August 12, 2020
- Ended
- April 22, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.