Device recall · October 2, 2019
Medfusion¿ Model 4000 Syringe Infusion Pump recall
Why: Customers who utilize PharmGuard¿ Server Software (PGS) with Medfusion¿ 4000 pumps manufactured prior to November 10, 2017, may experience wireless network connectivity issues on certain pumps due to a potential Radio Module issue on some Medfusion¿ 4000 pumps.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Medfusion¿ Model 4000 Syringe Infusion Pump, with PharmGuard¿ Server Software (PGS). Sold under the following names: MEDFUSION¿ 4000 Pump V1.0; MEDFUSION¿ 4000 Pump V1.1, v1.1.1, v1.1.2; MEDFUSION¿ 4000 Pump V1.1, V1.1.1, V1.1.2; MEDFUSION¿ 4000 Pump V1.5, V1.5.1; MEDFUSION¿ 4000 Pump V1.5, V1.5.1; MEDFUSION¿ 4000 Pump V1.6, V1.6.1 Product Usage: The pumps are indicated for the following uses: 1. In the administration of fluids requiring precisely controlled infusion rates including blood or blood products, lipids, drugs, antibiotics, enteral solutions and other therapeutic fluids; 2. By the following delivery routes: arterial, epidural, intravenous, intrathecal, subcutaneous, and enteral; 3. By the following delivery modes: continuous, volume/time, mass, body weight, intermittent and bolus.
- Lots / codes
Model Numbers (Serial Numbers): a. 4000-0100-50 (2000245, 2000246, 2000247, 2000248, 2000251, 2000252, 2000254, 2000255, 2000256, 2000257, 2000258, 2000260, 2000261, 2000264, 2000265, 2000267, 2000268, 2000269, 2000270, 2000272, 2000273, 2000274, 2000276,2000277, 2000278, 2000279, 2000280, 2000281, 2000282, 2000283, 2000284, 2000285, 2000286, 2000287, 2000288, 2000290, 2000291, 2000297, 2000298, 2000299, 2000300, 2000302, 2000304, 2000308, 2000309, 2000310, 2000311, 2000312, 2000313, 2000314, 2000315, 2000317, 2000318, 2000320, 2000321, 2000322, 2000324, 2000325, 2000326, 2000329, 2000331, 2000332, 2000336, 2000338, 2000341, 2000343, 2000345, 2000346, 2000348, 2000349, 2000352, 2000353, 2000355, 2000356, 2000358, 2000359, 2000360, 2000363, 2000364, 2000366, 2000370, 2000371, 2000374, 2000375, 2000376, 2000377, 2000379, 2000380, 2000381, 2000383, 2000385, 2000386, 2000387, 2000388, 2000389, 2000390, 2000392, 2000393, 2000395, 2000396, 2000403, 2000408, 2000409, 2000413, 2000415, 2000417, 2000426, 2000433, 2000434, 2000305, 2000330, 2000333, 2000357, 2000361, 2000362, 2000365, 2000369, 2000382, 2000391, 2000397, 2000398, 2000400, 2000402, 2000405, 2000414, 2000419, 2000423, 2000425, 2000430, 2000431, 2000439, 2000440, 2000441, 2000443, 2000316, 2000428, 2000438, 2000262, 2000342, 2000263, 2000327, 2000334, 2000378, 2000303, 2000351, 2000437, 2000301, 2000249, 2000275, 2000292, 2000373, 2000407, 2000394, 2000295, 2000335, 2000435, 2000422, 2000293, 2000418, 2000259, 2000367, 2000253, 2000416, 2000289, 2000344, 2000372, 2000350, 2000266, 2000250, 2000328, 2000401, 2000384, 2000404, 2000319, 2000399, 2000339, 2000412, 2000294, 2000340, 2000296, 2000429, 2000337, 2000368, 2000323, 2000307, 2000442); b. 4000-0101-51 (2002111, 2002121, 2002126, 2002127, 2002156, 2002171, 2002172, 2002221, 2002228, 2002241, 2002251, 2002253, 2002255, 2002263, 2002265, 2002272, 2002283, 2002291, 2002307, 2002318, 2002324, 2002336, 2002337, 2002339, 2002344, 2002353, 2002355, 2002356, 2002359, 2002195, 2002196, 2002197, 2002198, 2002199, 2002378, 2002379, 2004959, 2002407, 2002409, 2002410, 2002411, 2002412, 2002414, 2002415, 2002416, 2002417, 2002418, 2002419, 2002420, 2002421, 2002423, 2002427, 2002428, 2002429, 2002431, 2002432, 2002433, 2002437, 2002438, 2002439, 2002440, 2002441, 2002443, 2002445, 2002447, 2002448, 2002451, 2002452, 2002453, 2002454, 2002455, 2002456, 2002459, 2002460, 2002461, 2002462, 2002463, 2002464, 2002465, 2002466, 2002468, 2002469, 2002470, 2002472, 2002473, 2002474, 2002475, 2002478, 2002480, 2002483, 2002484, 2002485, 2002488, 2002489, 2002490, 2002491, 2002492, 2002494, 2002495, 2002496, 2002497, 2002498, 2002579, 2002600, 2002608, 2002616, 2002619, 2002630, 2002634, 2002643, 2002649, 2002705, 2002719, 2002722, 2002744, 2002771, 2002799, 2002811, 2002824, 2002826, 2002833, 2002835, 2002842, 2002844, 2002846, 2002857, 2004163, 2004164, 2004165, 2004167, 2004168, 2004169, 2004171, 2004172, 2004174, 2004176, 2004177, 2004178...
- Quantity
- 20,884 devices
- Company
- Smiths Medical ASD Inc., Minneapolis, MN
- Distribution
- Worldwide distribution - Nationwide in the states of AL, AR, AZ, CA, CN, CO, CT, DC, FL, GA, ID, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV and country of Canada.
- Initiated
- October 15, 2018 · Voluntary: Firm initiated
- Classified
- September 23, 2019
- Reported
- October 2, 2019
- Ended
- October 11, 2022
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.