Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor¿ Space Sy recall
Why: Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor¿ Space Syringe Pump, Infusion Pump, 1 each Component: N/A
- Lots / codes
Lot Code: Model No 8713080U; UDI-DI 4046963915308; All Serial Numbers
- Quantity
- 1663
- Company
- B Braun Medical Inc, Breinigsville, PA
- Distribution
- Domestic distribution nationwide. International distribution to Canada.
- Initiated
- June 14, 2024 · Voluntary: Firm initiated
- Classified
- July 25, 2024
- Reported
- July 31, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.