Device recall · September 3, 2025
Spectrum IQ Infusion Pump recall
Why: Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Spectrum IQ Infusion Pump, Product Code 3570009
- Lots / codes
UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
- Quantity
- 6 units
- Company
- Baxter Healthcare Corporation, Deerfield, IL
- Distribution
- US distribution to states of: AZ, FL
- Initiated
- July 14, 2025 · Voluntary: Firm initiated
- Classified
- August 22, 2025
- Reported
- September 3, 2025
- How notified
- Telephone