Device recall · August 28, 2019
Microdot Xtra blood glucose test strips recall
Why: The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Microdot Xtra blood glucose test strips
- Lots / codes
LOT NUMBERS 7091101, 7091201, 7091801, 7091902, 7100402, 7100403, 7101002, 7101602, 7101702, 7102403, 7103002, 7110602, 7110703, and 7112702.
- Quantity
- 17,784 (USA)
- Company
- Cambridge Sensors Limited, Huntingdon, N/A
- Distribution
- Illinois
- Initiated
- July 26, 2018 · Voluntary: Firm initiated
- Classified
- August 22, 2019
- Reported
- August 28, 2019
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.