Device recall · August 20, 2025
Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40 recall
Why: Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Description: The Insightra Ultra IABP Catheter Kit is a 7 French catheter available with either a 40, 35, 30, 25, or 20 cc balloon. The catheter has two lumens and is designed to provide counterpulsation cardiac assist therapy. The outer lumen is a channel for helium used to inflate and deflate the balloon. The inner lumen is used for a guidewire and blood pressure measurement. It is packaged as a convenience kit with accessories (one-way valve, 50 cc syringe, guide wires, introducer needle, dilator, pressure tubing, Datascope & Arrow adaptors). The balloon inflation and deflation cycle is synchronized (using a commercially available control console) with the ECG or arterial pressure to provide counterpulsation synchronized with the heartbeat. It is intended to increase coronary perfusion, decrease the workload of the left ventricle and allow healing of the myocardium Component: No
- Lots / codes
Lot code: All lots / UDI: IMU7F-40: 00850012832546; IMU7F-35: 00850012832515; IMU7F-30: 00850012832485; IMU7F-25: None; IMU7F-20: None
- Quantity
- 24227 units
- Company
- Insightra Medical Inc, Clarksville, TN
- Distribution
- Worldwide - US Nationwide distribution in the states of TX, MO, FL, MI and the countries of Albania, Azerbaijan, Bangladesh, Brazil, Cambodia, Colombia, Hong Kong, India, Indonesia, Iran, Israel, Saudi Arabia, Kuwait, Lebanon, Malaysia, Nepal, Oman, Pakistan, Pakistan, Peru, Romania, Russia, Serbia, Singapore, United Arab Emirates.
- Initiated
- July 19, 2025 · Voluntary: Firm initiated
- Classified
- August 13, 2025
- Reported
- August 20, 2025
- Ended
- March 18, 2026
- How notified
- N/A
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.