Device recall · September 1, 2021
Monoject 0.9% Sodium Chloride Flush Syringe recall
Why: Potential for the plunger to draw back after the air has been expelled and reintroduce air back into the syringe.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
Monoject 0.9% Sodium Chloride Flush Syringe, 5mL Fill, STERILE, Product Code 8881570125 The proposed device is indicated for use in flushing compatible intravenous tubing systems and indwelling intravascular access devices.
- Lots / codes
20D0614 21D0894 20D0624 20D0674 20D0684 20H1504 20H1514
- Quantity
- 267,217,860 eaches in total
- Company
- Cardinal Health, Mansfield, MA
- Distribution
- US Nationwide Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY.
- Initiated
- August 4, 2021 · Voluntary: Firm initiated
- Classified
- August 21, 2021
- Reported
- September 1, 2021
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.