Device recall · May 20, 2026
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application recall
Why: Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump
- Lots / codes
Software application (app) version 1.x and 2.x, loaded on the following units: 1. Model Number CT900A, GTIN 00643169890923, Serial Numbers: NPL0004445; 2. Model Number CT900B, GTIN 00643169890947, Serial Numbers: NPL1000042, NPL1000045, NPL1000046, NPL1000048, NPL1000050, NPL1000051, NPL1000053, NPL1000054, NPL1000059, NPL1000060, NPL1000062, NPL1000065, NPL1000069, NPL1000070, NPL1000071, NPL1000072, NPL1000073, NPL1000074, NPL1000075, NPL1000077, NPL1000081, NPL1000082, NPL1000083, NPL1000085, NPL1000086, NPL1000087, NPL1000090, NPL1000091, NPL1000093, NPL1000094, NPL1000098, NPL1000117, NPL1000133, NPL1000135, NPL1000137, NPL1000138, NPL1000151, NPL1000159, NPL1000160, NPL1000164, NPL1000166, NPL1000167, NPL1000169, NPL1000173, NPL1000174, NPL1000179, NPL1000180, NPL1000183, NPL1000185, NPL1000186, NPL1000187, NPL1000188, NPL1000191, NPL1000192, NPL1000193, NPL1000194, NPL1000195, NPL1000196, NPL1000230, NPL1000235, NPL1000237, NPL1000242, NPL1000244, NPL1000245, NPL1000246, NPL1000248, NPL1000249, NPL1000252, NPL1000254, NPL1000257, NPL1000258, NPL1000281, NPL1000292, NPL1000297, NPL1000301, NPL1000303, NPL1000307, NPL1000321, NPL1000322, NPL1000323, NPL1000325, NPL1000326, NPL1000332, NPL1000334, NPL1000343, NPL1000361, NPL1000362, NPL1000363, NPL1000364, NPL1000365, NPL1000366, NPL1000367, NPL1000369, NPL1000370, NPL1000381, NPL1000483, NPL1000486, NPL1000487, NPL1000501, NPL1000502, NPL1000505, NPL1000506, NPL1000507, NPL1000509, NPL1000510, NPL1000512, NPL1000515, NPL1000516, NPL1000518, NPL1000519, NPL1000528, NPL1000529, NPL1000539, NPL1000540, NPL1000555, NPL1000556, NPL1000558, NPL1000560, NPL1000561, NPL1000562, NPL1000563, NPL1000564, NPL1000571, NPL1000573, NPL1000574, NPL1000575, NPL1000576, NPL1000577, NPL1000578, NPL1000580, NPL1000585, NPL1000592, NPL1000603, NPL1000606, NPL1000612, NPL1000627, NPL1000630, NPL1000631, NPL1000632, NPL1000634, NPL1000637, NPL1000641, NPL1000652, NPL1000653, NPL1001194, NPL1001195, NPL1001201, NPL1001204, NPL1001205, NPL1001206, NPL1001207, NPL1001208, NPL1001210, NPL1001211, NPL1001212, NPL1001216, NPL1001223, NPL1001224, NPL1001225, NPL1001226, NPL1001228, NPL1001229, NPL1001230, NPL1001231, NPL1001235, NPL1001236, NPL1001237, NPL1001238, NPL1001240, NPL1001241, NPL1001243, NPL1001244, NPL1001250, NPL1001251, NPL1001254, NPL1001259, NPL1001264, NPL1001265, NPL1001272, NPL1001273, NPL1001274, NPL1001277, NPL1001282, NPL1001283, NPL1001290, NPL1001291, NPL1001306, NPL1001307, NPL1001313, NPL1001314, NPL1001328, NPL1001330, NPL1001332, NPL1001333, NPL1001334, NPL1001336, NPL1001337, NPL1001339, NPL1001341, NPL1001345, NPL1001349, NPL1001350, NPL1001351, NPL1001353, NPL1001354, NPL1001360, NPL1001361, NPL1001362, NPL1001366, NPL1001367, NPL1001369, NPL1001372, NPL1001375, NPL1001380, NPL1001381, NPL1001385, NPL1001386, NPL1001389, NPL1001398, NPL1001407, NPL1001408, NPL1001410, NPL1001412, NPL1001419, NPL1001421, NPL1001423, NPL1001430, NPL1001431, NPL1001434, NPL1001435, NPL1001436...
- Quantity
- 10165 units
- Company
- Medtronic Neuromodulation, Minneapolis, MN
- Distribution
- Worldwide distribution.
- Initiated
- April 8, 2026 · Voluntary: Firm initiated
- Classified
- May 14, 2026
- Reported
- May 20, 2026
- How notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.