Device recall · August 6, 2025
SwabFlush Prefilled Syringe with ICU Medical SwabCaps recall
Why: SwabFlush syringes are affected by the ICU Medical recall of their SwabCap Disinfecting Cap due to a manufacturing issue which affects the seal between the foil lid and plastic container of the Swab Cap. The issue may result in a reduced level of disinfection of the luer access valve, potentially resulting in patient infection/sepsis.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
SwabFlush Prefilled Syringe with ICU Medical SwabCaps, 10mL Model no. EMZE010301
- Lots / codes
UDI-DI 00363807103016 (ea) 10363807103013 (box) 20363807103010 (case) Lots 3144609 3144857 3144886 3144967 3145090 3145186
- Quantity
- 1,427,100 eaches
- Company
- MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada and Panama.
- Initiated
- June 16, 2025 · Voluntary: Firm initiated
- Classified
- July 30, 2025
- Reported
- August 6, 2025
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.