Class ITerminatedZ-2165-2021
Device recall · August 18, 2021
Dose IQ Safety Software used with Spectrum IQ Infusion Pump recall
Why: Software issue: The defect creates a mismatch between linked drug identifiers (IDs) in the Dose IQ user interface (UI) and the binary drug library (BDL) loaded onto the Spectrum IQ pump.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
Dose IQ Safety Software used with Spectrum IQ Infusion Pump
- Lots / codes
Software version: All libraries initially created with version 9.0.x
- Quantity
- 61 units
- Company
- Baxter Healthcare Corporation, Deerfield, IL
- Distribution
- US Nationwide Distribution.
- Initiated
- July 7, 2021 · Voluntary: Firm initiated
- Classified
- August 11, 2021
- Reported
- August 18, 2021
- Ended
- June 20, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.