Device recall · June 10, 2020
Omnipod DASH Insulin Management System recall
Why: After the device has been in use for about 2 months, data processing in the PDM can be slowed such that the Bolus Calculator fails to accurately subtract the correct amout of IOB before suggesting a bolus amount.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Omnipod DASH Insulin Management System (mg/dL configuration), Catalog Number PT000010 M/D: INT1D001MG - Product Usage: is intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.
- Lots / codes
UDI: PT - 000010 - 10385082000139; Lot Numbers: L000202
- Quantity
- 400
- Company
- Insulet Corporation, Acton, MA
- Distribution
- The products were distributed to the following foreign countries: Italy, Netherlands, UK.
- Initiated
- April 2, 2020 · Voluntary: Firm initiated
- Classified
- May 29, 2020
- Reported
- June 10, 2020
- Ended
- September 14, 2021
- How notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.