The PCA 3 system cam be used in a wide range of clinical settings recall
Why: Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum, burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.
- Lots / codes
List No. 12384; Serial Numbers: 12920041 ; 12920048 ; 12920052 ; 12920056 ; 12920066 ; 12920069 ; 12920079 ; 12920082 ; 12920101 ; 12920102 ; 12920104 ; 12920105 ; 12920106 ; 12920107 ; 12920117 ; 12920121 ; 12920122 ; 12920152 ; 12920158 ; 12920159 ; 12920160 ; 12920174 ; 12920246 ; 12920248 ; 12920251 ; 12920252 ; 12920255 ; 12920257 ; 12920264 ; 12920265 ; 12920268 ; 12920270 ; 12920275 ; 12920276 ; 12920277 ; 12920278 ; 12920282 ; 12920287 ; 12920289 ; 12920291 ; 12920294 ; 12920304 ; 12920305 ; 12920311 ; 12920312 ; 12920320 ; 12920328 ; 12920330 ; 12920332 ; 12920340 ; 12920343 ; 12920346 ; 12920347 ; 12920352 ; 12920353 ; 12920357 ; 12920358 ; 12920359 ; 12920360 ; 12920362 ; 12920364 ; 12920369 ; 12920371 ; 12920372 ; 12920377 ; 12920381 ; 12920382 ; 12920383 ; 12920384 ; 12920387 ; 12920388 ; 12920390 ; 12920392 ; 12920398 ; 12920401 ; 12920403 ; 12920404 ; 12920405 ; 12920410 ; 12920412 ; 12920413 ; 12920415 ; 12920416 ; 12920420 ; 12920424 ; 12920425 ; 12920428 ; 12920429 ; 12920432 ; 12920433 ; 12920434 ; 12920435 ; 12920436 ; 12920441 ; 12920444 ; 12920446 ; 12920448 ; 12920449 ; 12920453 ; 12920455 ; 12920457 ; 12920458 ; 12920460 ; 12920461 ; 12920462 ; 12920466 ; 12920467 ; 12920468 ; 12920469 ; 12920472 ; 12920473 ; 12920476 ; 12920477 ; 12920478 ; 12920480 ; 12920481 ; 12920482 ; 12920484 ; 12920485 ; 12920486 ; 12920487 ; 12920488 ; 12920489 ; 12920490 ; 12920492 ; 12920496 ; 12920497 ; 12920498 ; 12920499 ; 12920501 ; 12920505 ; 12920506 ; 12920507 ; 12920508 ; 12920509 ; 12920510 ; 12920514 ; 12920516 ; 12920517 ; 12920518 ; 12920519 ; 12920520 ; 12920521 ; 12920522 ; 12920523 ; 12920524 ; 12920525 ; 12920526 ; 12920530 ; 12920533 ; 12920534 ; 12920535 ; 12920537 ; 12920540 ; 12920541 ; 12920542 ; 12920544 ; 12920545 ; 12920546 ; 12920548 ; 12920555 ; 12920558 ; 12920559 ; 12920560 ; 12920562 ; 12920565 ; 12920566 ; 12920567 ; 12920568 ; 12920569 ; 12920570 ; 12920571 ; 12920574 ; 12920576 ; 12920577 ; 12920578 ; 12920581 ; 12920582 ; 12920585 ; 12920586 ; 12920587 ; 12920589 ; 12920590 ; 12920591 ; 12920592 ; 12920593 ; 12920596 ; 12920598 ; 12920599 ; 12920600 ; 12920601 ; 12920602 ; 12920603 ; 12920604 ; 12920605 ; 12920606 ; 12920607 ; 12920608 ; 12920610 ; 12920611 ; 12920612 ; 12920614 ; 12920615 ; 12920620 ; 12920621 ; 12920623 ; 12920626 ; 12920627 ; 12920629 ; 12920632 ; 12920633 ; 12920634 ; 12920635 ; 12920639 ; 12920641 ; 12920644 ; 12920646 ; 12920647 ; 12920648 ; 12920649 ; 12920651 ; 12920654 ; 12920657 ; 12920659 ; 12920660 ; 12920663 ; 12920664 ; 12920665 ; 12920666 ; 12920668 ; 12920669 ; 12920670 ; 12920671 ; 12920674 ; 12920675 ; 12920677 ; 12920679 ; 12920680 ; 12920681 ; 12920682 ; 12920686 ; 12920688 ; 12920690 ; 12920691 ; 12920692 ; 12920696 ; 12920699...
- Quantity
- 8,216 infusion pump systems
- Company
- Hospira Inc., Lake Forest, IL
- Distribution
- Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:
- Initiated
- March 8, 2013 · Voluntary: Firm initiated
- Classified
- July 8, 2016
- Reported
- July 20, 2016
- Ended
- November 7, 2016
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.