Device recall · August 21, 2019
VOLUMAT MC AGILIA US recall
Why: The firm is correcting four software anomalies and KVO (keep vein open) end of infusion alarm priority in Software versions 1.7 and 1.9a of Volumat MC Agilia Software and versions 1.0 and 1.1 of Vigilant DrugLib
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
VOLUMAT MC AGILIA US, Infusion Pump, REF Z021135, software versions 1.7 and 1.9a Product Usage: The Volumat MC Agilia is a piece of transportable medical equipment intended to be used in healthcare facilities environment by healthcare trained professionals according to hospital protocols on adults, pediatrics and neonatal human patient to administer via a single channel or mounted on a multiple channels rack accessory: Intermittent or continuous delivery of parenteral fluids (solutions, colloids, parenteral nutrition) and medications (including but limited to diluted drugs, chemotherapy) through clinically accepted IV routes of administration). Transfusion of blood and blood derivatives products. Note that for transfusion procedures, use of a special administration set is required. Volumat MC Agilia can be used with the Dose Error Reduction Software (DERS) Vigilant Drug Lib to define the default and upper hard limits for the Bolus Volume, or dose, by drug. The health care professional can then operate the pump for that drug within the upper hard limits defined in the Drug Library.
- Lots / codes
All serial numbers distributed in the US
- Quantity
- 9461 units
- Company
- Fenwal Inc, Lake Zurich, IL
- Distribution
- US Nationwide Distribution
- Initiated
- June 24, 2019 · Voluntary: Firm initiated
- Classified
- August 12, 2019
- Reported
- August 21, 2019
- Ended
- March 6, 2023
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.