Device recall · July 2, 2025
Omnipod 5 Automated Insulin Delivery System recall
Why: Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all Product Description: The Omnipod 5 iOS App software is available to download through the Apple app store. The Omnipod 5 App is the user interface that controls the Omnipod 5 Automated Insulin Delivery System. It is used to activate/deactivate Pods, display alerts/alarms, and send insulin delivery commands for execution to the Pod. Only available in the US.
- Lots / codes
Model Number: PT-000664/M009-S-AP. UDI-DI: 10385083000336. All software versions 1.X.X
- Quantity
- 122,823 downloads
- Company
- Insulet Corporation, Acton, MA
- Distribution
- US market only- Although there are 122823 downloads of the application, our data shows there are only ~80k active users.
- Initiated
- May 23, 2025 · Voluntary: Firm initiated
- Classified
- June 24, 2025
- Reported
- July 2, 2025
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.