Device recall · June 25, 2025
ICU Medical Plum Solos recall
Why: Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
ICU Medical Plum Solos, Item number 400011001; infusion pump
- Lots / codes
UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173, 60000174, 60000176, 60000177, 60000178, 60000179, 60000180, 60000183, 60000187, 60000189, 60000190, 60000191, 60000192, 60000198, 60000199, 60000201, 60000203, 60000204, 60000205, 60000206, 60000207, 60000208, 60000209, 60000210, 60000212, 60000213, 60000214, 60000215, 60000216, 60000217, 60000223, 60000224, 60000225, 60000226, 60000227, 60000228, 60000229, 60000230, 60000234, 60000236, 60000237, 60000238, 60000239, 60000240, 60000241, 60000242, 60000244, 60000245, 60000246, 60000247, 60000249, 60000251, 60000252, 60000255, 60000257, 60000259, 60000263, 60000264, 60000265, 60000266, 60000268, 60000269, 60000271, 60000273, 60000274, 60000275, 60000276, 60000277, 60000279, 60000280, 60000281, 60000282, 60000283, 60000284, 60000285, 60000286, 60000288, 60000291, 60000292, 60000295, 60000296, 60000297, 60000298, 60000300, 60000155, 60000193, 60000197, 60000211, 60000243, 60000270, 60000272, 60000123, 60000168, 60000175, 60000195, 60000221, 60000256, 60000294.
- Quantity
- 214 units
- Company
- ICU Medical, Inc., Lake Forest, IL
- Distribution
- Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.
- Initiated
- April 22, 2025 · Voluntary: Firm initiated
- Classified
- June 18, 2025
- Reported
- June 25, 2025
- How notified
- Visit
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.