Drugs A-Z Is my price fair? Price changes Drug markups Drug shortages Recalls Medicare prices Upcoming generics What people paid Guides Share what you paid
Theme
Class IIOngoingZ-1964-2025

Device recall · June 18, 2025

Baxter Novum IQ Syringe Pump recall

Why: Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.

Class II: could cause temporary or reversible health problems; serious harm is remote.

The recall

Product

Baxter Novum IQ Syringe Pump, product code 40800BAXUS,

Lots / codes

UDI/DI 05413765852428, All Serial Numbers

Quantity
8027 units
Company
Baxter Healthcare Corporation, Deerfield, IL
Distribution
US Nationwide distribution.
Initiated
May 12, 2025 · Voluntary: Firm initiated
Classified
June 12, 2025
Reported
June 18, 2025
How notified
Letter

If you have this product

  1. Check the lot number on your package against the lots above. Only those lots are recalled.
  2. Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
  3. Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
  4. Return or dispose of recalled product as the company or your pharmacy advises.

Questions about this recall

Be kind and stay on topic. No personal health details, links or selling. This is not medical advice.

No comments yet. Questions, tips, what worked for you - start the conversation.

Help the next person

Know what your pharmacy pays.

Every price you share shows someone else what's fair - and which pharmacies charge the least. It takes 30 seconds and it's anonymous.