Device recall · June 18, 2025
Baxter Novum IQ Syringe Pump recall
Why: Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete, and the syringe is unloaded. A System Error 21502 (Flange Sensor Failure) may occur. Please note this issue only affects a subset of Novum IQ syringe pumps manufactured or repaired after July 4, 2023.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Baxter Novum IQ Syringe Pump, product code 40800BAXUS,
- Lots / codes
UDI/DI 05413765852428, All Serial Numbers
- Quantity
- 8027 units
- Company
- Baxter Healthcare Corporation, Deerfield, IL
- Distribution
- US Nationwide distribution.
- Initiated
- May 12, 2025 · Voluntary: Firm initiated
- Classified
- June 12, 2025
- Reported
- June 18, 2025
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.