Class IIOngoingZ-1937-2024
Device recall · June 12, 2024
MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE recall
Why: It has been identified that testing documentation does not support that the syringes can successfully deliver accurate volumes across the full range of their claimed graduated capacity and may pose a risk to patient health.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
MEDLINE LUER LOCK DISPOSABLE SYRINGE WITHOUT NEEDLE, REF SYR101010
- Lots / codes
UDI/DI 10884389596319, All lots in the format: 897YYMMXXXX
- Quantity
- 25,214,334 syringes
- Company
- MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Panama, UAE, USVI.
- Initiated
- April 22, 2024 · Voluntary: Firm initiated
- Classified
- May 31, 2024
- Reported
- June 12, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.