Device recall · June 16, 2021
JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" recall
Why: Specific models and lot numbers of the Jelco Hypodermic Needle-Pro Fixed Needle Insulin Syringe 29Gx1/2 odd number line graduation markings on the syringe barrel were skewed approximately 20 degrees upward.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
JELCO HYPODERMIC NEEDLE-PRO FIXED NEEDLE INSULIN SYRINGE: a) 1.0mL 28G X 1/2" (0.33mm x 13mm), REF 4428-1 b) 1.0mL 29G X 1/2" (0.33mm x 13mm), REF 4429-1
- Lots / codes
a) REF 4428-1, Lot Numbers: 4046543 and 4062235 b) REF 4429-1, Lot Numbers: 4031846, 4031845, 4040734, 4043536, 4046545, 4046546, 4062239, 4062240, 4062238, 4062242, 4014096
- Quantity
- 1,432,200 units
- Company
- Smiths Medical ASD Inc., Minneapolis, MN
- Distribution
- Worldwide distribution - US Nationwide and the countries of UK, Spain.
- Initiated
- May 17, 2021 · Voluntary: Firm initiated
- Classified
- June 10, 2021
- Reported
- June 16, 2021
- Ended
- October 11, 2023
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.