Device recall · April 5, 2017
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library recall
Why: SIGMA Spectrum Infusion Pumps with Master Drug Library, Software V8.00.02 contains an anomaly that may cause a Sharp Watchdog Timeout error message. Specifically, the software anomaly causes the error message only under conditions in which a message is being written to the Spectrum Infusion Pumps Event History Log (EHL), and the length of that message exactly matches the size of the available memory at the end of a memory sector. For this reason, the issue is significantly more likely to manifest when poor network maintenance results in high volumes of messages being written to the EHL.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library, Model #/Product Code 35700BAX2
- Lots / codes
14,362 Serial Numbers (please contact CDRH for list of affected Serial Numbers).
- Quantity
- Domestic: 14,175 units; Canada: 187 units
- Company
- Baxter Healthcare Corporation, Medina, NY
- Distribution
- Natiowide, Puerto Rico and Canada
- Initiated
- January 26, 2017 · Voluntary: Firm initiated
- Classified
- March 29, 2017
- Reported
- April 5, 2017
- Ended
- May 24, 2018
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).