Drugs A-Z Is my price fair? Price changes Drug markups Drug shortages Recalls Medicare prices Upcoming generics What people paid Guides Share what you paid
Theme
Class IITerminatedZ-1664-2018

Device recall · May 16, 2018

Baxter Sigma Spectrum Infusion Pump with master Drug Library recall

Why: Service technician improperly documented final release testing on product, indicating that the full battery of release tests were performed when, in fact, several tests were omitted. he omitted tests include air-in-line detection at 400ml/hr and 800ml/hr and verification of upstream occlusion detection at 800ml/hr. Improper detection of air-in-line or upstream occlusion may result in serious adverse health consequences.

Class II: could cause temporary or reversible health problems; serious harm is remote.

The recall

Product

Baxter Sigma Spectrum Infusion Pump with master Drug Library (Version 6) 35700BAXL Product Usage: The Baxter Sigma Spectrum Infusion Pump with master Drug Library is intended to be used for the controlled administration of fluids. These fluids may includes pharmaceutical drugs, blood, blood products and mixture if required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluids. Spectrum infusion pump is intended to be used in conjunction with legally marketed intravenous administration sets and medications provided by user. The Spectrum and Spectrum with MAster Drug Library is intended to reduce operator interaction through automated programming thereby helping to reduce errors associated with complex device programming.

Lots / codes

Serial Numbers 709860 735355 743891 749663 753665 757109 763909 768077 774906 776397 780931 785500 789177 791625 792667 794258 797835 798697 799183 801072 802788 803224 804175 804202 804235 804561 804988 806824 811741 812442 812989 813180 844637 844720 845844 846093 846115 858309 864606 864957 870921 872713 872941 873294 873296 873302 877416 879360 883764 885021 905402 910300 912661 914647 915558 921852 923011 923362 923483 923990 926408 926680 927374 927490 927810 927848 928705 929190 929572 929887 930694 930914 930972 931613 931745 932170 932315 932496 932601 933050 933625 933745 934174 934407 934629 934656 934684 934685 935474 935554 935636 936378 936422 937342 938176 938742 939342 939364 939571 942194 945100 962029 965017 965310 967527 982596 983087 996433 996840 997035 997317 999136 1000594 1006058 1006935 1008672 1011288 1011482 1012217 1012228 1012708 1012883 1013017 1013180 1013371 1013628 1015173 1015537 1015796 1015902 1016229 1016277 1016287 1017064 1017616 1017627

Quantity
444 units in total
Company
Baxter Healthcare Corporation, Medina, NY
Distribution
US Nationwide & Puerto Rico
Initiated
April 3, 2018 · Voluntary: Firm initiated
Classified
May 4, 2018
Reported
May 16, 2018
Ended
February 14, 2020
How notified
Letter

If you have this product

  1. Check the lot number on your package against the lots above. Only those lots are recalled.
  2. Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
  3. Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
  4. Return or dispose of recalled product as the company or your pharmacy advises.

Recalled with it (same recall event)

Questions about this recall

Be kind and stay on topic. No personal health details, links or selling. This is not medical advice.

No comments yet. Questions, tips, what worked for you - start the conversation.

Help the next person

Know what your pharmacy pays.

Every price you share shows someone else what's fair - and which pharmacies charge the least. It takes 30 seconds and it's anonymous.