Device recall · May 15, 2024
t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology recall
Why: During normal use, the mobile app version 2.7 may crash and be automatically relaunched by the iOS operating system. This cycle intermittently repeats, which leads to excessive Bluetooth communication that may result in pump battery drain and may lead to the pump shutting down sooner than typically expected. Pump shutdown will cause insulin delivery to suspend, which could lead to an under-delivery of insulin and may result in hyperglycemia, including severe hyperglycemia.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
t:connect mobile app used in conjunction with t:slim X2 insulin pump with Control-IQ technology
- Lots / codes
version 2.7 or later /UDI: 00850006613274
- Quantity
- 85,863 applications
- Company
- Tandem Diabetes Care, Inc., San Diego, CA
- Distribution
- US nationwide distribution including Puerto Rico.
- Initiated
- March 5, 2024 · Voluntary: Firm initiated
- Classified
- May 6, 2024
- Reported
- May 15, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall