Device recall · April 24, 2024
Smiths Medical Medfusion Model 3500 Syringe pumps recall
Why: Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Smiths Medical Medfusion Model 3500 Syringe pumps, Item Numbers: a) 3500 b) 3500-0600-00 c) 3500-0600-01 d) 3500-0600-249 e) 3500-0600-50 f) 3500-0600-51 g) 3500-0600-82 h) 3500-306 i) 3500-402 j) 3500-414 k) 3500-415 l) 3500-500 m) 3500E n) 3500VX-306 o) 3500VX-500
- Lots / codes
All serial numbers Medfusion Model 3500 pumps manufactured or serviced with these potentially affected Barrel Clamp Guides between Jul-2016 and Apr-2021. a) 3500, UDI/DI NONE, Serial Numbers: M02939, M03874, M09235, M09277, M09440, M09611, M09616, M10171, M10178, M10184, M10211, M10277, M10385, M10621, M11229, M12020, M12026, M12252, M12253, M12254, M13692, M14074, M14737, M15127, M15867, M15889, M16087, M16525, M17236, M17252, M18256, M18273, M18495, M18544, M18589, M18697, M18871, M19131, M21028, M21729, M21739, M22438, M24137, M24448, M28363, M28500, M28504, M30668, M30820, M32347, M36928, M39766; b) 3500-0600-00, UDI/DI 10610586041259, Serial Numbers: M107413, M107445, M107557, M107606, M107707, M107718, M110082, M110289, M110363, M111064, M73212, M-8833, M-8834, M-9284, M-9285, M-9286, M-9301, M-9302, M-9303; c) 3500-0600-01, UDI/DI 10610586041242, Serial Numbers: M100101, M102639, M102877, M103027, M103063, M103064, M103065, M103068, M103093, M103095, M103097, M103098, M103099, M103100, M103101, M103191, M103193, M103203, M103230, M103231, M103232, M103288, M103308, M103309, M103467, M103597, M103598, M103599, M103605, M103608, M103632, M103655, M103656, M103703, M103704, M103705, M103706, M103721, M103777, M103780, M103781, M103852, M103910, M103912, M103913, M103914, M103915, M103917, M103918, M103919, M103920, M103923, M103939, M103965, M103968, M104023, M104031, M104077, M104081, M104082, M104086, M104167, M104168, M104196, M104203, M104214, M104215, M104216, M104224, M104225, M104226, M104227, M104240, M104247, M104250, M104335, M104336, M104340, M104342, M104573, M104574, M104586, M105017, M105388, M105406, M105427, M105494, M105523, M105539, M105546, M105550, M105560, M105563, M105638, M105654, M105661, M105682, M105830, M105893, M105955, M106092, M106254, M106370, M106398, M106403, M106404, M106433, M106457, M106460, M106463, M106506, M106522, M106600, M106604, M106623, M106627, M106628, M106631, M106632, M106659, M106677, M106681, M106715, M106718, M106724, M106727, M106729, M106732, M106738, M106739, M106743, M106744, M106747, M106750, M106756, M106769, M106773, M106774, M106782, M106786, M106790, M106910, M106911, M106925, M106937, M106938, M106964, M106978, M106992, M106993, M106994, M106995, M106996, M106998, M107000, M107003, M107009, M107010, M107011, M107012, M107014, M107016, M107022, M107104, M107105, M107106, M107107, M107108, M107109, M107110, M107111, M107112, M107113, M107114, M107115, M107116, M107117, M107118, M107119, M107120, M107121, M107122, M107123, M107124, M107125, M107126, M107127, M107128, M107129, M107130, M107131, M107132, M107133, M107134, M107135, M107136, M107137, M107138, M107139, M107140, M107141, M107142, M107143, M107144, M107145, M107146, M107147, M107148, M107149, M107150, M107151, M107152, M107153, M107154, M107155, M107156, M107157, M107158, M107159, M107160, M107161, M107162, M107163, M107164, M107165, M107166, M107167, M107168, M107169, M107170, M107171, M107172, M107173, M107174...
- Quantity
- 3821 pumps
- Company
- Smiths Medical ASD Inc., Minneapolis, MN
- Distribution
- Worldwide distribution.
- Initiated
- December 19, 2023 · Voluntary: Firm initiated
- Classified
- April 15, 2024
- Reported
- April 24, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.