Class IITerminatedZ-1520-2017
Device recall · March 29, 2017
Alaris System PC unit recall
Why: Reports where the Low Battery alarm and/or the Very Low Battery alarm are not being triggered before the battery is discharged and all infusion channels are stopped.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Alaris System PC unit, model no. 8000 and 8015. The central programming, monitoring and power supply component for the Alaris infusion pump System.
- Lots / codes
all units
- Quantity
- 613,800 total units (575,221 units in US)
- Company
- CareFusion 303, Inc., San Diego, CA
- Distribution
- Worldwide distribution. US nationwide, Europe, Australia, New Zealand, South Africa, Greater Asia, Middle East, and Canada.
- Initiated
- November 1, 2016 · Voluntary: Firm initiated
- Classified
- March 21, 2017
- Reported
- March 29, 2017
- Ended
- April 24, 2020
- How notified
- Letter