Device recall · April 23, 2025
t:slim X2 Insulin Pump with Interoperable Technology recall
Why: A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
t:slim X2 Insulin Pump with Interoperable Technology
- Lots / codes
Software versions 7.9/UDI: 00389152407012
- Quantity
- 63 units
- Company
- Tandem Diabetes Care, Inc., San Diego, CA
- Distribution
- US: AZ, CA, CO, CT, FL, HI, IA, ID, IL, IN, MD, ME, MI, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, PA, TX, UT, VA, WA, WI OUS: N/A
- Initiated
- February 27, 2025 · Voluntary: Firm initiated
- Classified
- April 11, 2025
- Reported
- April 23, 2025
- How notified
- Telephone
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.