Device recall · April 22, 2015
Master Drug Library Software version 8.0 recall
Why: Loading/Bolus default dose settings in the Master Drug Library and the values shown on the pump during programming may differ. MDL drug dose time in seconds will round to the nearest integer in minutes on the pump dose setup screen (20 sec may show as 1 min on the pump display). The pump will administer drugs as configured. The discrepancy may cause therapy delay or unintended rate of delivery.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Master Drug Library Software version 8.0, Product Code 35723V080, to be used with SIGMA Spectrum Infusion System (Pump) version 8.0, Product Code 35700BAX2 The SIGMA Spectrum Infusion Pump with Master Drug Library (MDL) is intended to be used for the controlled administration of fluids.
- Lots / codes
Software version 8.0, Product Code 35723V080
- Quantity
- 58 MDLs
- Company
- Baxter Healthcare Corp, Deerfield, IL
- Distribution
- Worldwide Distribution: US Distribution to states of: CA, CT, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NH, NY, SD, TX, UT, WV, including Puerto Rico and country of: Canada.
- Initiated
- March 2, 2015 · Voluntary: Firm initiated
- Classified
- April 16, 2015
- Reported
- April 22, 2015
- Ended
- May 22, 2018
- How notified
- Telephone
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.