Device recall · March 22, 2017
CADD Solis VIP Ambulatory Infusion Pump recall
Why: 20 Pumps sold to the Finnish market contain a message in which one word in the message is mistranslated. When the user follows a specific set of key presses the pump will display the incorrect message. The message indicates that a Patient Controlled Analgesia (PCA) dose is unavailable because the pump is running. It should indicate that the PCA dose is not available because the pump is stopped. The function of the pump is unchanged and no patient injury can occur since no drug is being delivered.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
CADD Solis VIP Ambulatory Infusion Pump, Model 21-21210, Reorder 21-2120-0102-15,
- Lots / codes
1071155, 1076826, 1081258, 1081260, 1081275, 1082475, 1082511, 1082512, 1090347, 1090348, 1090349, 1091767, 1091768, 1091769, 1091770, 1091771, 1092395, 1100186, 1100188, 1100189.
- Quantity
- 20
- Company
- Smiths Medical ASD, Inc., Saint Paul, MN
- Distribution
- Internationally to Finland
- Initiated
- October 31, 2016 · Voluntary: Firm initiated
- Classified
- March 10, 2017
- Reported
- March 22, 2017
- Ended
- July 15, 2019
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall