Device recall · July 20, 2022
MiniMed 620G Insulin Pump recall
Why: Due to battery cap deterioration, this may potentially result in an incomplete battery circuit (similar to when battery cap is not attached) and a loss of power/functionality of the insulin infusion pumps.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
MiniMed 620G Insulin Pump (O.U.S. version)
- Lots / codes
Product Number/CFN (O.U.S. version) MiniMed 620G (MMT-1510, MMT-1710, MMT-1750) UDI Code: Pending Serial Numbers: NG1003917H NG1004461H NG1010799H NG1012557H NG1020680H NG1026110H NG1028941H NG1028949H NG1028952H NG1028973H NG1034230H NG1034247H NG1034304H NG1034314H NG1034335H NG1034338H NG1034345H NG1034358H NG1034359H NG1034364H NG1034583H NG1034584H NG1034593H NG1034595H NG1034609H NG1035490H NG1035495H NG1035505H NG1035517H NG1035520H NG1035525H NG1035526H NG1035546H NG1035552H NG1035566H NG1035572H NG1035575H NG1035581H NG1035586H NG1035593H NG1035595H NG1035915H NG1035930H NG1035933H NG1035935H NG1036645H NG1036646H NG1036671H NG1036678H NG1036688H NG1036704H NG1036713H NG1036715H NG1036725H NG1036726H NG1036731H NG1036757H NG1036785H NG1036801H NG1036809H NG1036811H NG1036813H NG1036854H NG1039152H NG1040219H NG1040230H NG1040231H NG1040237H NG1040257H NG1040262H NG1040271H NG1040290H NG1040298H NG1040305H NG1040318H NG1040344H NG1040353H NG1040380H NG1040381H NG1040383H NG1040480H NG1040483H NG1040628H NG1040637H NG1040667H NG1040668H NG1040737H NG1040999H NG1041013H NG1041041H NG1041067H NG1041078H NG1041079H NG1041134H NG1041165H NG1041169H NG1041212H NG1041229H NG1041254H NG1041262H NG1043526H NG1043542H NG1043547H NG1043549H NG1043570H NG1043571H NG1043574H NG1043585H NG1043587H NG1043589H NG1043600H NG1043605H NG1043618H NG1043619H NG1043629H NG1043637H NG1043642H NG1043656H NG1043659H NG1048215H NG1051985H NG1051990H NG1052098H NG1054459H NG1054461H NG1055929H NG1055937H NG1055959H NG1055980H NG1055987H NG1056018H NG1056021H NG1056025H NG1057684H NG1057686H NG1057690H NG1059069H NG1059076H NG1059083H NG1059107H NG1059119H NG1059125H NG1059126H NG1059151H NG1059154H NG1059178H NG1059187H NG1059297H NG1059307H NG1059313H NG1059319H NG1059326H NG1059356H NG1059367H NG1059386H NG1059411H NG1059419H NG1059424H NG1059440H NG1060999H NG1066735H NG1066747H NG1066778H NG1066783H NG1066788H NG1066803H NG1066816H NG1066821H NG1066823H NG1066827H NG1066854H NG1066863H NG1066871H NG1066891H NG1066893H NG1066894H NG1066897H NG1066899H NG1066902H NG1066904H NG1066906H NG1066909H NG1066916H NG1068481H NG1068815H NG1068819H NG1068829H NG1068831H NG1068832H NG1068838H NG1068841H NG1068842H NG1068850H NG1068892H NG1073307H NG1073320H NG1073321H NG1073336H NG1073354H NG1073371H NG1073381H NG1073392H NG1073406H NG1073412H NG1073413H NG1073415H NG1076044H NG1076047H NG1076063H NG1076079H NG1076086H NG1076103H NG1076128H NG1076132H NG1076148H NG1076153H NG1076156H NG1076170H NG1076181H NG1076187H NG1076212H NG1076217H NG1082514H NG1082516H NG1082524H NG1082527H NG1082532H NG1082534H NG1082538H NG1082559H NG1082564H NG1082567H NG1082568H NG1082571H NG1082572H NG1082587H NG1082589H NG1082594H NG1082595H NG1082597H NG1082602H NG1082641H NG1082645H NG1082650H NG1082667H NG1082672H NG1082680H NG1082682H NG1082684H NG1082694H NG1082700H NG1082702H NG1082705H NG1082737H NG1082739H NG1082741H NG1084306H NG1084330H NG1084341H NG1084348H NG1084350H NG1084360H...
- Quantity
- 2,287 pumps
- Company
- Medtronic MiniMed, Northridge, CA
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, and WY O.U.S.: Not provided
- Initiated
- May 23, 2022 · Voluntary: Firm initiated
- Classified
- July 8, 2022
- Reported
- July 20, 2022
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.