Device recall · March 12, 2025
Philips Connected Blood Glucose Meter recall
Why: Telcare, LLC has received customer complaints related to battery expansion/swelling for its Philips Connected Blood Glucose Meter (i.e., BGM 4 brand name) Monitoring System.. The issue was identified by post market surveillance data (customer complaints) - events of battery expansion/swelling and associated symptoms, e.g., leading to the device's removable back cover (battery port) to separate. If there is a loss of primary function of the device due to the described device defect, an immediate health consequence that may result is prolonged hypoglycemia or hyperglycemia due to delay in treatment, as the blood glucose measurement is not available to guide treatment. In a worst-case situation, there is potential for battery swelling to result in explosion of the battery and/ or device. Exposure to explosion, and subsequently projectiles from the explosion, may result in several immediate health consequences, including concussion, headache, dizziness, eye irritation, blurred vision, eye injury, and vision impairment.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Philips Connected Blood Glucose Meter, BGM 4; Model Number: TM0009;
- Lots / codes
Model Number: TM0009; UDI-DI: 00859519002407; Serial Numbers: Serial Numbers are formatted as TFXXXXXX or TGXXXXXX (X represents digits). Meter Serial Numbers: TF048802, TF098191, TF079123, TF076786, TF076836, TF089080, TF088984, TF086097, TF085390, TF085029, TF084238, TF098751, TF098975, TF098976, TF096704, TF096961, TF096966, TF117862, TF127077, TF049583, TF047153, TF063135, TF127267, TF100376, TF086287, TF104396, TF104399, TF104401, TF104403, TF104404, TF104405, TF104407, TF104408, TF104409, TF104410, TF104411, TF104412, TF104414, TF104415, TF104416, TF104417, TF104418, TF104421, TF104426, TF104427, TF104428, TF104429, TF104430, TF104431, TF104432, TF104433, TF104439, TF104440, TF104441, TF104442, TF104444, TF104446, TF104448, TF104449, TF104452, TF104453, TF104454, TF104455, TF104456, TF104458, TF104460, TF104461, TF104462, TF104467, TF104468, TF104469, TF104471, TF104473, TF104474, TF104479, TF104480, TF104481, TF104482, TF104483, TF104486, TF104487, TF104490, TF104491, TF104492, TF104494, TF104496, TF104498, TF104500, TF104501, TF104503, TF104505, TF104506, TF104510, TF104511, TF104512, TF104514, TF104516, TF104520, TF104521, TF104522, TF104525, TF104526, TF104527, TF104528, TF104529, TF104530, TF104534, TF104535, TF104537, TF104538, TF104541, TF104545, TF104554, TF104555, TF104566, TF104567, TF104568, TF104570, TF104571, TF104573, TF104576, TF104577, TF104579, TF104581, TF104585, TF104586, TF104588, TF104589, TF104590, TF104592, TF104593, TF104595, TF104599, TF104601, TF104602, TF104605, TF104609, TF104610, TF104615, TF104618, TF104620, TF104621, TF104625, TF104626, TF104628, TF104629, TF104630, TF104632, TF104634, TF104635, TF104639, TF104641, TF104642, TF104643, TF104644, TF104646, TF104652, TF104657, TF104660, TF104661, TF104662, TF104663, TF104664, TF104666, TF104667, TF104670, TF104673, TF104674, TF104676, TF104677, TF104680, TF104681, TF104684, TF104693, TF104694, TF104695, TF104697, TF104699, TF104716, TF104718, TF104721, TF104724, TF104725, TF104726, TF104727, TF104729, TF104733, TF104735, TF104739, TF104740, TF104742, TF104745, TF104746, TF104747, TF104749, TF104751, TF104753, TF104754, TF104760, TF104761, TF104762, TF104770, TF104771, TF104777, TF104782, TF104789, TF104790, TF104791, TF104794, TF104795, TF111281, TF111284, TF111290, TF111291, TF111292, TF111295, TF111297, TF111304, TF111305, TF111307, TF111310, TF111312, TF111317, TF111329, TF111330, TF111331, TF111334, TF111335, TF111344, TF111345, TF111346, TF111355, TF111356, TF111357, TF111359, TF111364, TF111366, TF111369, TF111370, TF111371, TF111375, TF111379, TF111380, TF111383, TF111388, TF111389, TF111390, TF111392, TF111393, TF111427, TF111438, TF111445, TF111448, TF111455, TF111458, TF111469, TF111471, TF111472, TF111486, TF111494, TF111498, TF111499, TF111500, TF111501, TF111503, TF111504, TF111516, TF111517, TF111519, TF111520, TF111524, TF111528, TF111530, TF111534, TF111536, TF111538, TF111539, TF111540, TF111542, TF111545, TF111548, TF111550, TF111597, TF111598...
- Quantity
- 34,764 units
- Company
- Telcare, LLC, Concord, MA
- Distribution
- US Nationwide distribution.
- Initiated
- January 30, 2025 · Voluntary: Firm initiated
- Classified
- March 5, 2025
- Reported
- March 12, 2025
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall