Device recall · June 22, 2022
t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM recall
Why: Insulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the battery life drops to less than 5%, insulin delivery will continue for 30 minutes and then the pump will power off and insulin delivery will stop. If the battery reaches 1%, then insulin delivery will stop, which could lead to hyperglycemia.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
t:slim X2 Insulin Pump with Dexcom G5 Mobile CGM, and t:slim X2 Insulin Pump with Basal-IQ Technology
- Lots / codes
Dexcom G5 - All Serial Numbers; Basal-IQ Technology - Software Versions 6.3.01 and Earlier
- Quantity
- 5,206
- Company
- Tandem Diabetes Care Inc, San Diego, CA
- Distribution
- International distribution in the country of Australia.
- Initiated
- September 10, 2020 · Voluntary: Firm initiated
- Classified
- June 10, 2022
- Reported
- June 22, 2022
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.