Device recall · March 17, 2021
Omnipod DASH Insulin Management System User Guide recall
Why: Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Omnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX XX PT 000318 PT 000319 Distributed to: Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland
- Lots / codes
Part Number/Rev./Country: PT 000002 AUT DEU mg 1, 2, 3 Guide, Dash User, German Austria PT 000002 BEL DEU MG 1, 2, 3 Guide, Dash User Guide, Belgium Germany, mgdl PT 000002 BEL FRE MG 1, 2, 3, 4 Guide, Dash User Guide, Belgium French, mgdl PT 000002 BEL NLD MG 1, 2, 3 Guide, Dash User Guide, Belgium Dutch, mgdl PT 000002 DEU DEU mg 1, 2, 3 Guide, Dash User Guide, German Germany, mgdl PT 000002 DEU DEU mm 1, 2, 3 Guide, Dash User Guide, German Germany, mmoll PT 000002 DNK DANmm 1, 2, 3 Guide, Dash User Guide, Danish Denmark PT 000002 FIN FIN mm 1, 2, 3 Guide, Dash User Guide, Finnish Finland PT 000002 FRA FRE mg 1, 2, 3, 4 Guide, Dash User Guide, French France PT 000002 NLD NLD mm 1, 2, 3, 4 Guide, Dash User Guide, Dutch Netherland PT 000002 NOR NORmm 1, 2, 3 Guide, Dash User Guide, Norwegian Norway PT 000002 SWE SWEmm 1, 2, 3 Guide, Dash User Guide, Swedish Sweden 41333 CAN 1, 2 User Guide, Omnipod, Dash, Canada PT 000318 1, 2 Guide, Dash User Guide, Greece PT 000319 1, 2 Guide, Dash User Guide, Turkey
- Quantity
- 21,723 guides
- Company
- Insulet Corporation, Acton, MA
- Distribution
- Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland
- Initiated
- February 1, 2021 · Voluntary: Firm initiated
- Classified
- March 10, 2021
- Reported
- March 17, 2021
- Ended
- February 8, 2023
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.