Device recall · February 19, 2020
HUDSON RCI COMFORT FLO Corrugated Humidification System recall
Why: There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
HUDSON RCI COMFORT FLO Corrugated Humidification System, REF 2415, humidifier nebulizer kit
- Lots / codes
GTIN 14026704659361 Lot Numbers: 74D1501285 74J1500185 74M1500350 74A1601241 74C1601016 74C1601472 74C1602180 74D1600257 74D1601233 74D1601232 74E1600054 74G1600014 74H1603490 74J1602886 74J1602885 74B1701597 74F1700031 74G1700590 74G1702197 74J1700026 74J1701902 74K1700026 74K1700900 74L1700541 74A1800863 74A1802331 74B1801424 74C1800022 74D1801520 74E1800781 74G1800923 74K1800769 74M1800651 74A1902597 74D1900087 74D1901071
- Quantity
- 30566 devices
- Company
- Teleflex Medical Inc, Morrisville, NC
- Distribution
- nationwide
- Initiated
- December 13, 2019 · Voluntary: Firm initiated
- Classified
- February 7, 2020
- Reported
- February 19, 2020
- Ended
- April 30, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.