Device recall · February 11, 2015
Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabete recall
Why: Calibration factors in the pump overwritten during a programming step. The force sensor could send a lower signal value to the pump processor, with loss of prime warnings, occlusion alarms and the pump unable to detect a cartridge during the prime sequence. Field action initiated 8/29/2011.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Animas Vibe Insulin Infusion Pump. This product is indicated for continuous subcutaneous infusion of insulin for the treatment of diabetes and has a continuous glucose monitoring feature.
- Lots / codes
Model Number(s): 100515-63 100510-63 100514-63 100512-63 100511-63 101200-03 100201-03 101202-03 101204-03 101205-03 101200-53 101202-53 101205-53 101200-02 101200-57 101201-57 101202-57 101204-57 101205-57 101200-63 101201-63 101202-63 101204-63 101205-63 101206-63
- Quantity
- 1235
- Company
- Animas Corporation, West Chester, PA
- Distribution
- No US distribution, Distributors are located in France, Germany, Sweden and United Kingdom.
- Initiated
- September 6, 2011 · Voluntary: Firm initiated
- Classified
- January 30, 2015
- Reported
- February 11, 2015
- Ended
- August 10, 2015
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.