Device recall · March 21, 2018
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library recall
Why: Certain infusion pumps may exhibit excessive wear of the pumping mechanism caused by a lack of lubrication, this could affect the occlusion of the tubing and result in over- or under-infusion.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 6) (GTIN 00085412091570); Model Numbers: 35700BAX and 35700ABB. Intended to be used for the controlled administration of fluids.
- Lots / codes
700000 700001 700008 400012 700014 700015 700038 400040 400044 400057 700060 400068 700070 400073 400076 700078 700079 400085 700088 700099 700105 400106 400114 700117 700131 700137 400141 700148 700180 700194 700197 400226 700227 700228 700229 700232 700236 700239 400245 700252 700254 700260 700262 700274 400282 700291 700300 700305 700306 400311 700313 700336 700347 700351 700352 400355 700356 700358 400368 700374 700376 700379 400382 700398 700400 700412 700434 700447 700465 700467 700470 400474 400480 700482 700495 700500 700517 700521 700526 700535 700544 700548 700555 700574 700585 400592 700610 400620 400621 400634 700668 700670 700671 700680 400682 700689 700691 400747 400748 400752 700757 400761 700762 700778 700802 700814 700820 400828 400844 700850 700863 700872 700883 700887 700888 700895 700901 700906 700929 700940 400948 400984 701013 401017 701021 401028 701041 401082 701084 401091 701123 701136 701176 401177 401186 401209 701210 401220 701221 701224 701232 701277 701312 401313 701327 401341 701358 701362 701382 701451 701455 701458 701459 701463 701466 701469 701474 701476 701478 701489 701490 701497 701498 701500 701503 701509 701512 701521 701522 701523 701526 701544 401666 401677 701710 701736 401751 701772 701778 701816 701821 401830 701833 401844 401849 401850 401851 401856 701857 401860 701864 701868 401874 701877 401889 401891 701894 401903 401907 401931 701935 401938 701944 401946 401957 701960 401970 401978 701984 701990 701991 702000 702002 702003 702005 702017 702018 702020 702021 702022 702030 702032 702033 702037 702042 702043 702044 702046 702049 702051 702063 702064 702066 702071 702073 702077 402097 702103 402108 702112 702114 402116 702129 702130 702145 702149 702179 402196 402203 702263 702322 702354 702386 702444 702447 702473 702474 702483 702503 702506 702538 702549 702592 402601 702624 702649 702669 702673 702677 702679 702692 702701 702703 702705 702728 702729 402761 702784 702785 702787 702789 702799 702802 702804 702805 702811 702813 702816 702818 702824 702828 702835 702836 702838 702839 702846 702847 702848 702851 702857 702858 702860 702862 702866 702867 702869 702870 702872 702877 702878 702880 702886 702891 702906 702909 702914 402945 702955 702964 702981 702983 702985 702986 702992 702993 702994 702998 703001 703002 703005 703010 703014 703019 703021 703022 703025 703026 703029 703030 703035 703036 703037 703039 703044 703046 703049 703053 703060 703074 703077 703090 703092 703096 703097 703102 703107 703108 703113 703114 703116 703118 703119 703128 703129 703131 703133 703137 703138 703141 703151 703154 703157 703158 703159 703160 703161 703163 703164 703165 703167 703168 703170 703171 703172 703177 703181 703184 703189 703190 703207 703208 703209 703214 703216 703221 703223 703231 703236 703238 703240 703241 703242 703243 703245 703246 703251 703254 703255 703266 703268 703273 703285 703289 703292 703294 703297 703307 703309 703311 703315 703318 703321 703322 703324...
- Quantity
- 173,375
- Company
- Baxter Healthcare Corporation, Medina, NY
- Distribution
- Worldwide Distribution - US, including the District of Columbia, Puerto Rico, and Canada.
- Initiated
- October 6, 2017 · Voluntary: Firm initiated
- Classified
- March 12, 2018
- Reported
- March 21, 2018
- Ended
- April 11, 2019
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.