Device recall · December 17, 2025
MEDLINE PERFUSION SYRINGE recall
Why: Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing BD Luer Tip Caps. BD has issued a recall of their Luer Tip Caps after confirming through internal testing that certain BD Luer Tip Cap trays failed routine biocompatibility testing. Although the tip cap covers a small surface area of the syringe and does not come in direct contact with patients, there is the potential that the failure of a biocompatibility test may have varying consequences ranging from no harm or clinical impact to potential for sensitization, irritation, pyrogenicity, acute systemic toxicity or hemolysis depending on the nature and duration of contact with the syringe.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
MEDLINE PERFUSION SYRINGE, REF: DYNJ39223C
- Lots / codes
UDI/DI 10193489621532 (each) 40193489621533 (case), Lot Number 25FMJ081
- Quantity
- 15 units
- Company
- Medline Industries, LP, Northfield, IL
- Distribution
- US Nationwide distribution in the state of CA.
- Initiated
- October 27, 2025 · Voluntary: Firm initiated
- Classified
- December 5, 2025
- Reported
- December 17, 2025
- How notified
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.