Drugs A-Z Is my price fair? Price changes Drug markups Drug shortages Recalls Medicare prices Upcoming generics What people paid Guides Share what you paid
Theme
Class IITerminatedZ-0852-2016

Device recall · March 2, 2016

AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant recall

Why: There were two potential interference conditions identified with the way the tubing set attaches to the inserter.

Class II: could cause temporary or reversible health problems; serious harm is remote.

The recall

Product

AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.

Lots / codes

Catalog number 900301, Lot #01131543, 01261508, 10151404, 10271408, 12161403, 12161404, 09021404, 01261509, 10271407, 01261507, 10071415, 05011403, 06241407, 08071402, 10091305, 03041403, 01271412, 07281401, and 11061303. Catalog number 905103, 01131542, 11241408, 12161405, 12161406, 0211501, 01211502, 11241406, 02031403, 06241408, 07161407, 09101402, 09281401, 10251303, 05301402, 11241407, and 12101305.

Quantity
413 units
Company
Stryker Spine, Allendale, NJ
Distribution
US Nationwide Distribution
Initiated
December 16, 2015 · Voluntary: Firm initiated
Classified
February 23, 2016
Reported
March 2, 2016
Ended
August 1, 2016
How notified
Letter

If you have this product

  1. Check the lot number on your package against the lots above. Only those lots are recalled.
  2. Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
  3. Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
  4. Return or dispose of recalled product as the company or your pharmacy advises.

Questions about this recall

Be kind and stay on topic. No personal health details, links or selling. This is not medical advice.

No comments yet. Questions, tips, what worked for you - start the conversation.

Help the next person

Know what your pharmacy pays.

Every price you share shows someone else what's fair - and which pharmacies charge the least. It takes 30 seconds and it's anonymous.