Device recall · February 14, 2024
Cardinal Health Monoject recall
Why: The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.
Class I: a reasonable chance the product could cause serious harm or death.
The recall
- Product
Cardinal Health Monoject, 6ML Enteral Syringe (purple) Low Dose (LDT) - For Use with ENFit Connection System, REF 406SE
- Lots / codes
UDI/DI 26971564466220 (Case), 16971564466223 (Box), 06971564466226 (Each), Lot Numbers: 230503 and 230701
- Quantity
- 20186 units
- Company
- Cardinal Health 200, LLC, Waukegan, IL
- Distribution
- Distribution US nationwide and Canada. Foreign distribution updated 9/6/2024 to only include Canada.
- Initiated
- December 28, 2023 · Voluntary: Firm initiated
- Classified
- February 2, 2024
- Reported
- February 14, 2024
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.