Class IIOngoingZ-0776-2025
Device recall · January 8, 2025
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library recall
Why: Devices were identified as released after repair without full testing being performed, which includes flow testing.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product Code 35700BAX2
- Lots / codes
GTIN 00085412498683; Serial Number: 2153638
- Quantity
- 1 unit
- Company
- Baxter Healthcare Corporation, Deerfield, IL
- Distribution
- US Nationwide distribution in the states of AL, IN.
- Initiated
- November 19, 2024 · Voluntary: Firm initiated
- Classified
- December 27, 2024
- Reported
- January 8, 2025
- How notified
- Telephone
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.