Device recall · December 30, 2020
30" (76 cm) Appx 3.3 ml recall
Why: Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
30" (76 cm) Appx 3.3 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Hanger, REF: CL3011; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL3011T; 40" (102 cm) Appx 4.9 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Drip Chamber, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3020; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Hanger, Drop-In Red Cap, REF: CL3028; 30" (76 cm) Appx 6.1 ml, 20 Drop Admin Set w/Integrated ChemoLock Drip Chamber, 0.2 Micron Filter, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3028T; 42" (107 cm) Appx 5.2 ml, Admin Set w/20 Drop In-Line Chamber, ChemoLock w/Red Cap, Spiros, Bag Hanger, Drop-In Red Cap, REF: CL3361; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock Additive Port, Vented Cap; ChemoLock; Spinning Spiros w/Red Cap, REF: CL3538; Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros; ChemoLock; ChemoLock Universal Vented Vial Spike; Bag Hanger, REF CL3929; 8.5" (22 cm) Appx 1.0 ml, Ext Set w/MicroClave Clear, ChemoLock Port, Spiros w/Red Cap, Clamp, REF: CL3931; 7.5" (19 cm) Appx 1.1 ml, Bifuse Ext Set w/ChemoLock Port, ChemoClave, Spiros, 2 Clamps, REF: CL3932; 30 IN(76cm)APPX 2.1ml, EXT SET w/ChemoLock w/RED CAP, Spiros RED CAP, REF: CL3935; Oncology Kit w/ChemoLock w/ChemoClave; Spinning Spiros w/ Red Cap, REF: CL3941; Oncology Kit w/5" (13 cm) Add-On Set w/ChemoLock w/Red Cap, Vented Cap; ChemoLock Port w/Bag Spike; Spiros w/Red Cap, REF: CL3946; Oncology Kit, ChemoLock Bag Spike w/Additive Port, Vented Cap, Spiros, REF: CL3948; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960; Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros, ChemoLock, Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960T; 60" (152 cm) Appx 7.4 ml, Admin Set w/ChemoLock w/Red Cap, 20 Drop In-Line Chamber, Spiros w/Red Cap, Bag Hanger, REF: CL4100; 30" (76 cm) Appx 3.6 ml, Admin Set w/20 Drop In-Line Drip Chamber, ChemoLock Additive Port, Spiros w/Red Cap, REF: CL4131; 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer, REF: CL4151; 31" (79 cm) Appx 3.4 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, Spiros w/Red Cap, Drop-In ChemoLock Injector, Bag Hanger, REF: CL3978
- Lots / codes
Lots: 4720479, 4724743, 4735699, 4775051, 4803047, 4808098, 4868079, 4868080, 4735696, 4763384, 4800367, 4866943, 4908482, 4724700, 4742639, 4758414, 4775030, 4787228, 4808060, 4848204, 4712385, 4724701, 4605106, 4733557, 4757782, 4775514, 4800370, 4724728, 4765922, 4867983, 4724703, 4749489, 4803140, 4878873, 4742641, 4758385, 4794501, 4720364, 4752843, 4799894, 4850394, 4866135, 4790865, 4735669, 4742630, 4758441, 4764898, 4787232, 4848213, 4867995, 4743901, 4765927, 4775070, 4777973, 4794503, 4606697, 4735662, 4803124, 4808069, 4867961, 4897098, 4898098, 4899830, 4573189, 4606684, 4749454, 4756268, 4803136, 4808059, 4848185, 4878880, 4726940, 4742091, 4738028, 4749474, 4560357, 4735675, 4714539, 4763384, 4800367, 4866943, 4908482, 4742641
- Quantity
- 231250
- Company
- ICU Medical, Inc., San Clemente, CA
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of FL, MD, LA, PA, IN, TX, NY, MO, VA, OH, GA, NC, IL, CA, AZ, AK, NE, DE, WI, CO, UT, ND, MS, MA, MN, WA, SC, OR, KS, MI, NH, NJ, KY, TN, ID, AR, CT, OK, AL, SD, ME, IA, VT, WY, DC, NV, RI, MT, NM, HI, WV and the countries of Italy, Belgium, United Kingdom, Canada, Germany, Argentina, Australia, Poland, Greece, Switzerland, Israel, France, Finland, Spain, Malaysia, Taiwan, Ireland, Luxembourg, Hong Kong , Brazil, Netherlands, Austria, Sweden, Turkey, Singapore, Slovakia, Croatia, South Africa, Romania, Thailand.
- Initiated
- August 31, 2020 · Voluntary: Firm initiated
- Classified
- December 23, 2020
- Reported
- December 30, 2020
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Recalled with it (same recall event)
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.