Device recall · December 14, 2022
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM recall
Why: The firm has become aware of an issue with the Omnipod 5 Controller where the Controller charging port and charging cable are melting, deforming, or discoloring due to heat generated by increased resistance in the Controller port. The excess heat may cause minor burns if touched, and the Omnipod 5 Controller may fail to charge, requiring the user to switch to their backup insulin plan.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Omnipod 5 Automated Insulin Delivery System Product catalog numbers: PT-000409: ASM, Omnipod 5, PDM, Programmed PT-000428: Adaptor, Charging, Noetic Nuu N5004L, Horizon PDM, US PT-000429: Cable, Charging, Omnipod 5, Horizon PDM
- Lots / codes
ALL serial numbers/ controller lots are currently in scope of this recall
- Quantity
- 129,086
- Company
- Insulet Corporation, Acton, MA
- Distribution
- Domestic distribution nationwide.
- Initiated
- November 14, 2022 · Voluntary: Firm initiated
- Classified
- December 7, 2022
- Reported
- December 14, 2022
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.