Drugs A-Z Is my price fair? Price changes Drug markups Drug shortages Recalls Medicare prices Upcoming generics What people paid Guides Share what you paid
Theme
Class IIOngoingZ-0265-2027

Device recall · October 28, 2026

Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 recall

Why: A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is an optional feature that, when enabled, triggers a notification at the scheduled time to remind users to take their long-acting insulin dose. The issue was caused by a type cast exception during database migration that occurs upon app update to version 1.2.1 causing a silent failure that prevented the creation of new alerts. Failure to receive the long-acting dose reminder may result in missed or delayed insulin doses, potentially leading to hyperglycemia. Conversely, if users are uncertain whether they took their dose due to the missing reminder, inadvertent duplicate dosing may occur, potentially resulting in hypoglycemia.

Class II: could cause temporary or reversible health problems; serious harm is remote.

The recall

Product

Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components: " Simplera sensor " Instinct sensor " InPen smart insulin pen " Apple Watch [Optional] Continuous glucose monitoring (CGM) is a technology that allows you to continuously view glucose values. The app uses an inserted Simplera sensor CGM or Instinct sensor CGM with Bluetooth¿ wireless technology that converts small amounts of glucose from the interstitial fluid under the skin into an electronic signal. The sensors use that signal to provide sensor glucose (SG) values to the MiniMed Go app on a compatible mobile device. The app also provides alerts based on your SG levels and contains a user interface for entering data such as insulin, meals, exercise, blood glucose (BG) values, and uploading information to the CareLink" Personal website. The app is available to download from the Apple"* App Store"* or the Google Play"* Store. Component: No

Lots / codes

Lot Code: GTIN for MMT-8600, iOS: 00199150038978 GTIN for MMT-8601, Android: 00199150038985 UDI for MMT-8600, iOS: 00199150038978 UDI for MMT-8601, Android: 00199150038985 Software version 1.2.1

Quantity
623
Company
Medtronic MiniMed, Inc., Northridge, CA
Distribution
Worldwide - U.S. Nationwide distribution in the states of AL, CA, CT, FL, LA, MA, MD, NY, OH, PA, and TX. The countries of Austria, Belgium, Denmark, Finland, Germany, Iceland, Italy, Norway, Spain, Sweden, Switzerland, United Kingdom.
Initiated
September 2, 2026 · Voluntary: Firm initiated
Classified
October 22, 2026
Reported
October 28, 2026
How notified
Letter

If you have this product

  1. Check the lot number on your package against the lots above. Only those lots are recalled.
  2. Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
  3. Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
  4. Return or dispose of recalled product as the company or your pharmacy advises.

Questions about this recall

Be kind and stay on topic. No personal health details, links or selling. This is not medical advice.

No comments yet. Questions, tips, what worked for you - start the conversation.

Help the next person

Know what your pharmacy pays.

Every price you share shows someone else what's fair - and which pharmacies charge the least. It takes 30 seconds and it's anonymous.