Device recall · November 18, 2015
Triton Infusion Pump recall
Why: The device might fail to detect air in line while infusing the medication Venofer. This may occur when a drop of Venofer gets stuck in the tubing adjacent to the bubble detector and the Total Volume Limit is programmed higher than the actual bag volume.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Triton Infusion Pump (model 300000) and Triton fp Infusion Pump (model 400000) (proprietary name: AFF (Mark I) Volumetric Infusion Pump). Used to pump fluids into a patient in a controlled manner.
- Lots / codes
Model numbers 300000, 400000
- Quantity
- 4695 units
- Company
- WalkMed Infusion, LLC, Englewood, CO
- Distribution
- Distributed in the states of AL, CA, CO, DE, FL, GA, IL, MA, MD, MI, MN, MS, NC, NJ, NM, NV, NY, OH, PA, TN, TX, and WA.
- Initiated
- October 1, 2015 · Voluntary: Firm initiated
- Classified
- November 6, 2015
- Reported
- November 18, 2015
- Ended
- July 12, 2016
- How notified
- Letter
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.