Device recall · November 11, 2015
Helioseal F Sealant Refill 1 x 1.25 g recall
Why: A defect occurred in the production process of Helioseal F Sealant; the particles in the faulty component seem to be larger than the size defined in the formula. This results in difficulties during extrusion of the material.
Class II: could cause temporary or reversible health problems; serious harm is remote.
The recall
- Product
Helioseal F Sealant Refill 1 x 1.25 g, REF 558519AN, and Helioseal F Assortment (kit) 5 x 1.25 g, REF 558518AN. Dental sealant. -- ivoclar vivadent clinical --- Helioseal F Refill is provided in a syringe packed sealed foil bag. Helioseal F is a light-curing, white-shaded fissure sealant featuring fluoride release. Helioseal fissure sealants are suitable for the sealing of pits, fissures and foramina caeca.
- Lots / codes
Lot U12511, Kit Lot U19925 (Helioseal F Assortment only), exp. 11.09.2017
- Quantity
- Domestic: 337 units
- Company
- Ivoclar A. G., Schaan, N/A
- Distribution
- Nationwide Distribution including NJ, PA, SC, FL, IN, IA, WI, IL, MO, TX, CA and WA.
- Initiated
- October 7, 2015 · Voluntary: Firm initiated
- Classified
- November 4, 2015
- Reported
- November 11, 2015
- Ended
- June 14, 2016
- How notified
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
If you have this product
- Check the lot number on your package against the lots above. Only those lots are recalled.
- Don't stop a prescription medicine on your own. Call your pharmacy - they can swap it for an unaffected lot or another maker.
- Felt unwell? Contact your doctor, and report it to FDA MedWatch (1-800-FDA-1088 or fda.gov/medwatch).
- Return or dispose of recalled product as the company or your pharmacy advises.
Questions about this recall
No comments yet. Questions, tips, what worked for you - start the conversation.